Exploratory evaluation of negative-pressure operating modes for size- and position-specific aerosol containment during simulated cardiopulmonary resuscitation in a newly developed respiratory isolation chamber

J.H. Suh1, J.Y. Hong1, K.S. Sung2
Available online: October 22, 2026
1Department of Emergency Medicine, Yonsei University College of Medicine, Seoul, Korea
2SS-ENG Co., Ltd., Bucheon, Korea
Corresponding author:  J.Y. Hong,
Email: juyoungbaby@yuhs.ac
Received: 20 May 2026   • Revised: 14 August 2026   • Accepted: 14 August 2026
  • 229 Views
  • 3 Download
  • 0 Crossref
  • 0 Scopus

Objective
To compare negative-pressure operating modes for aerosol containment during simulated cardiopulmonary resuscitation (CPR) in a newly developed double-barrier respiratory isolation chamber, focusing on particle-size– and operator-position–specific particle-count patterns.
Methods
In this mannequin-based benchtop simulation, a standardized nebulized aerosol challenge was tested under four modes: 0 Pa, −10 Pa, nominal “Al Pa” adaptive pressure control, and −50 Pa. Particle counts across six size channels were measured at intubator and chest-compressor positions, with and without CPR. Each chamber-applied condition was assessed in 10 re-initialized runs. Analyses were descriptive and estimation-based, using absolute counts, percentage reductions relative to corresponding positive controls, and direct mode comparisons.
Results
During CPR, residual particle counts differed by particle size and operator position. Under 0 Pa at the chest-compressor position, mean counts were 199,724.3, 79,870.1, and 12,580.0 for 0.3, 0.5, and 1.0 μm particles; corresponding Al Pa counts were 17,539.3, 1,826.1, and 222.3. Under Al Pa, mean reductions relative to corresponding positive controls ranged from 92.2% to >99.9% across CPR submicron comparisons. Compared with −50 Pa, Al Pa showed 11.1%–30.2% lower mean residual submicron counts.
Conclusion
During simulated CPR, aerosol containment varied by particle size and operator position. Thresholdbased adaptive pressure modulation maintained low residual submicron counts without continuous maximal suction and may offer a balanced strategy by applying additional suction only when needed. Patient-side safety and airflow effects require further evaluation.

Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:

Include:

Exploratory evaluation of negative-pressure operating modes for size- and position-specific aerosol containment during simulated cardiopulmonary resuscitation in a newly developed respiratory isolation chamber
Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:
Include:
Exploratory evaluation of negative-pressure operating modes for size- and position-specific aerosol containment during simulated cardiopulmonary resuscitation in a newly developed respiratory isolation chamber
Close